PAT700 Total Organic Carbon Analyzer – Diagnostics Page 129
Anatel Operator Manual
Figure 8-8 Plumbing test dialog box
Flow path selection drop down list box allows you to select from 12 pre-configured valve
and pump combinations. You may select none (default), bypass, flow through cell, flow
through cell with pump, sample from bottle 1, sample from bottle 2, sample from bottle 3,
sample from bottle 4, back flush through bottle 1, back flush through bottle 2, back flush
through bottle 3, or back flush through bottle 4.
None. Selecting this option turns off the pump and closes all the valves.
Bypass. Selecting this option opens only the bypass valve.
Flow through cell. Selecting this option opens only the cell valve.
Flow through cell with pump. Selecting this option opens the cell valve, the pump
valve, and turns on the pump.
Sample from bottle 1, bottle 2, bottle 3, or bottle 4. Selecting this option opens the
selected bottle valve, the pump valve, and turns on the pump.
Back flush through bottle 1, bottle 2, bottle 3, or bottle 4. Selecting this option
opens the selected bottle valve and the cell valve.
Return icon returns you to the diagnostics dialog box.
Load bottles icon opens the select bottles to load dialog box. See Figure 8-9.
Unload bottles icon takes you to the unload/replace bottles dialog box. See Figure 8-11,
page 131.