Universal Security Instruments MP118 Manual de usuario Pagina 134

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 194
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 133
Page 126 Diagnostics – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Figure 8-4 Diagnostics dialog box, test tab
RS-232 test
The RS-232 test icon allows you to test serial communication from the serial port by using the
printer port as a return port. See Figure 8-5.
1) Disconnect the printer and RS-232 serial connections and connect a serial cable
between the analyzer’s serial ports (from one serial port to the other).
2) Press the start test icon.
Figure 8-5 RS-232 test dialog box
Start test icon sends serial data out one serial port and reads the data on the other. If the
read data matches the sent data, the test passes.
Return icon returns you to the diagnostics dialog box.
Vista de pagina 133
1 2 ... 129 130 131 132 133 134 135 136 137 138 139 ... 193 194

Comentarios a estos manuales

Sin comentarios